|
Autor:
Dean R. Hess, PhD, RRT
Associate
Professor of Anesthesia, Harvard
Medical School
Assistant
Director of Respiratory Care,
Massachusetts General Hospital
Boston, MA USA |
|
|
NEW MODELS OF MECHANICAL VENTILATION
|
Introduction
Positive pressure ventilation
is described by trigger, control, and cycle
variables. The trigger initiates a breath.
The control variable remains constant
throughout inspiration regardless of changes
in respiratory system impedance. Inspiration
ends when the cycle is reached. The
relationship between the various possible
breath types and conditional variables is
the mode of ventilation. A mode can include
pressure and volume controlled breaths and
can be sophisticated enough to switch from
one control variable to the other. With each
generation of ventilators, new modes and
other features become available. The purpose
of this paper is to describe the technical
aspects of new modes and related features of
mechanical ventilators that recently have
become available.
The Ventilator Trigger
Patient triggering is usually
pressure-tirggered or flow-triggered.
Pressure triggering requires sufficient
patient inspiratory effort to cause airway
pressure to fall from the set end-expiratory
level to a threshold level (sensitivity) set
by the clinician. With flow triggering,
breath initiation is based on a flow change
in the ventilator circuit beyond some
pre-determined threshold. From the available
evidence, the following recommendations can
be made. 1) The trigger on current
generation ventilators is superior to that
which existed in the past. Auto-triggering
can be problematic secondary to the artifact
(e.g., cardiac oscillations) when the
trigger is very sensitive. 2) There is no
clear superiority of flow triggering and
pressure triggering. The choice of trigger
type should be based on patient response,
using the trigger type that produces the
best patient comfort. 3) Patient difficulty
triggering the ventilator is usually due to
pathophysiology (e.g., auto-PEEP) rather
than the trigger type or sensitivity.
Pressure Ventilation
Pressure controlled
ventilation
With pressure-controlled
ventilation (PCV), the ventilator applies a
set pressure to the airway for a set
inspiratory time. Pressure-controlled
breaths can be either patient-triggered or
ventilator-triggered. Tidal volume during
PCV is determined a number of variables: the
pressure control setting, airways
resistance, respiratory system compliance,
auto-PEEP, and patient effort. Inspiratory
time also affects the tidal volume if the
flow does not decrease to zero. Inspiratory
flow is fixed with volume-controlled
ventilation (VCV), whereas flow with PCV is
variable. Because of this, PCV may be
desirable to VCV if the patient is
triggering the ventilator with a strong
respiratory drive. PCV has been advocated by
some authorities as a lung protective
strategy and to improve patient-ventilator
synchrony. However, it has recently been
shown that, with lung protective
ventilation, the work of breathing is
greater with pressure-controlled ventilation
(compared to volume-controlled ventilation).
More important, with pressure controlled
ventilation, volume and trans-pulmonary
pressure limitation is not assured if the
patient makes vigorous inspiratory efforts.
Pressure-controlled
inverse-ratio ventilation
Early reports of improved
oxygenation with pressure-controlled
inverse-ratio ventilation (PCIRV) generated
considerable enthusiasm for this method.
Following the initial enthusiasm for this
approach, a subsequent controlled studies
reported no benefit or marginal benefit for
the use of PCIRV. Based on the available
evidence, there seems to be no clear role
for PCIRV in the management of patients with
ARDS. The likelihood of an improvement in
oxygenation using inverse ratio ventilation
is small and the risk of auto-PEEP and
hemodynamic compromise is great.
Pressure support ventilation
Pressure support ventilation
(PSV) assists inspiratory muscles during
invasive and noninvasive ventilation. PSV is
patient triggered and primarily flow
triggered. Secondary cycling mechanisms with
PSV are pressure and time. Thus, PSV cycles
to the expiratory phase when the flow
decelerates to a ventilator-determined
level, when the pressure rises to a
ventilator-determined level, or the
inspiratory time reaches a
ventilator-determined limit. Although PSV is
often considered a simple mode of
ventilation, in reality it can be quite
complex: 1) the ventilator must recognize
the patient?s inspiratory effort, which
depends on the trigger sensitivity of the
ventilator and the presence of auto-PEEP. 2)
The ventilator must deliver an appropriate
flow at the onset of inspiration. A flow
that is too high can produce a pressure
overshoot, whereas a flow that is too low
can produce patient flow starvation and
dyssynchrony. 3) The ventilator must
appropriately cycle to the expiratory phase
without the need for active exhalation by
the patient.
Like PCV, the flow
deceleration during PSV is largely a
function of the resistance and compliance of
the respiratory system. The flow at which
the ventilator cycles can either a fixed
absolute flow, a flow based on the peak
inspiratory flow, or a flow based on peak
inspiratory flow and elapsed inspiratory
time. Several studies have reported
dyssynchrony with PSV in subjects having
airflow obstruction (e.g., COPD). With
airflow obstruction, the inspiratory flow
decelerates slowly during PSV, the flow
necessary to cycle may not be reached, and
this stimulates active exhalation to
pressure cycle the breath. This problem
increases with higher levels of PSV and with
higher levels of airflow obstruction.
Several approaches can be used to solve this
problem. 1) PCV can be used, with the
inspiratory time set short enough so that
the patient does not contract the expiratory
muscles to terminate inspiration. 2) On some
newer generation ventilators, the clinician
can adjust the termination flow at which the
ventilator cycles.
The flow at the onset of the
inspiratory phase is determined by rise time
- the time required for the ventilator to
reach the PSV level at the onset of
inspiration. Newer generation of ventilators
allow adjustments of the rise time during
PSV. The rise time is adjusted to patient
comfort and ventilator graphics may be
useful to guide this setting. In patients
with a strong respiratory drive, a rapid
rise time may decrease the work of breathing
and the patient?s sensation of dyspnea.
However, patient comfort may be compromised
using rise times that are either to low or
too high. Moreover, a high inspiratory flow
at the onset of inspiration is not
necessarily beneficial for several reasons.
First, if the flow is higher at the onset of
inspiration, the inspiratory phase may be
prematurely terminated if the ventilator
cycles to the expiratory phase at a flow
that is a fraction of the peak inspiratory
flow. Second, the existence of a
flow-related inspiratory terminating reflex
in the airway has recently been described.
Activation of this reflex due to a higher
inspiratory flow causes shortening of neural
inspiration, which could result in brief,
shallow inspiratory efforts.
Another issue with PSV is the
presence of leaks in the system (e.g.,
bronchopleural fistula, cuffless airway,
mask leak with noninvasive ventilation). If
the leak exceeds the termination flow at
which the ventilator cycles, either active
exhalation will occur to terminate
inspiration or a prolonged inspiratory time
will be applied. With a leak, either PCV or
a ventilator that allows an adjustable
termination flow should be used.
Proportional Assist
Ventilation
Proportional assist
ventilation (PAV) was designed to increase
or decrease airway pressure in proportion to
patient effort, which should improve
patient-ventilator synchrony. This is
accomplished by a positive feedback control
that amplifies airway pressure
proportionally to inspiratory flow and
volume, where respiratory elastance and
resistance are the feedback signal gains.
Unlike other modes of ventilatory support,
which deliver a preset tidal volume or
inspiratory pressure at the airway, with
proportional assist ventilation the amount
of support changes with patient effort,
assisting ventilation with a uniform
proportionality between ventilator and
patient. The advantage of a proportional
ventilatory support lies in its ability to
track changes in ventilatory effort. To the
extent that inspiratory effort is a
reflection of ventilatory demand, this form
of support may result in a more physiologic
breathing pattern.
Tube Compensation
Tube compensation (TC) is
designed to compensate for endotracheal tube
resistance via closed loop control of
calculated tracheal pressure. The proposed
advantages of ATC are to overcome the
work-of-breathing imposed by artificial
airways, to improve patient/ventilator
synchrony as a result of variable
inspiratory flow commensurate with demand,
and to reduce air-trapping as a result of
compensation for imposed expiratory
resistance. This system uses the known
resistive coefficients of the tracheal tube
(tracheostomy or endotracheal) and
measurement of instantaneous flow to apply
pressure proportional to resistance
throughout the total respiratory cycle.
Because in vivo tracheal tube resistance
tends to be greater than in vitro
resistance, incomplete compensation for
endotracheal tube resistance may occur.
Additionally, kinks or bends in the tube as
it traverses the upper airway and
accumulation of secretions in the inner
lumen will change the tube?s resistive
coefficient and result in incomplete
compensation.
Whether endotracheal tube
resistance poses a clinical concern for
increased work-of-breathing in adults is
controversial. The imposed work-of-breathing
through the endotracheal tube is modest at
usual minute ventilations for the tube sizes
most commonly used for adults. Several
recent studies cast doubt on the importance
of endotracheal tube resistance during short
trials of spontaneous breathing. For
example, similar outcomes have been reported
when spontaneous breathing trials were
conducted with PSV (7 cm H2O) or
with a T-piece. Moreover, it has been
reported that the work-of-breathing through
the endotracheal tube amounted to only about
10% of the total work-of-breathing. The
work-for-breathing during a two-hour
spontaneous breathing trial with a T-piece
may be similar to the work-of-breathing
immediately following extubation. Although
prolonged spontaneous breathing through an
endotracheal tube is not desirable due to
the resistance of the tube, this may not be
important for short periods of spontaneous
breathing to assess extubation readiness.
Airway pressure-release
ventilation
Airway pressure-release
ventilation (APRV) produces alveolar
ventilation as an adjunct to continuous
positive airway pressure (CPAP). Airway
pressure is transiently released to a lower
level, after which it is quickly restored to
reinflate the lungs. Because the patient is
allowed to breathe spontaneously at both
levels of CPAP, the need for sedation is
potentially decreased, hemodynamics are
potentially better, dependent atelectasis
may be less, and oxygenation may be better.
Tidal volume for the APRV breath depends on
lung compliance, airways resistance, the
magnitude of the pressure release, the
duration of the pressure release, and the
magnitude of the patient?s spontaneous
breathing efforts. Of concern is the
potential for alveolar derecruitment during
the release of pressure with APRV.
A modification of APRV is the
situation in which the I:E ratio is not
reversed. This is available on some
ventilators as PCV+ (called BIPAP in Europe)
or Bilevel. Without spontaneous breathing,
PCV+ is similar to PCV and APRV is similar
to PCIRV. One potential advantage of these
modes is that the exhalation valve is active
during both the inspiratory and expiratory
phase. Prior to the current generation of
ventilators, the exhalation valve was active
during the expiratory phase, but closed
completely during the inspiratory phase. An
active exhalation valve during the
inspiratory phase will open as necessary to
maintain a constant inspiratory pressure.
The use of APRV has become fashionable in
some trauma centers, but evidence is lacking
for improved patient outcomes compared to
traditional ventilator modes.
One use of PCV+ (or Bilevel)
is to provide sighs during PCV or CPAP. With
this technique, several periods (2 to 4/min)
of elevated airway pressure (25 to 35 cm H2O)
is used periodically as a sigh (1 to 3
seconds at the higher pressure level). This
approach differs from the sighs that were
available in older generation of ventilators
in several ways. 1) They are provided more
frequently. 2) They are pressure limited. 3)
They are applied for a time longer than the
typical inspiratory times set on the
ventilator. 4) Due to the active exhalation
valve, the patient can continue to breathe
spontaneously at the higher pressure.
Although this strategy is attractive in
spontaneously breathing patients prone to
develop atelectasis, its benefit to date is
anecdotal.
Dual Control Modes
Recently developed modes
allow the ventilator to control pressure or
volume based on a feedback loop (dual
control). It is important to appreciate,
however, that the ventilator can only
pressure or volume - not both at the same
time. Dual control within a breath describes
a mode where the ventilator switches from
pressure control to volume control during
the breath. Dual control breath-to-breath is
simpler because the ventilator operates in
the either PCV or PSV, and the pressure
limit increases or decreases to maintain the
selected tidal volume.
Dual Control Breath-to-breath
? Pressure limited flow cycled ventilation
Breath-to-breath dual control
is available on several ventilators as
Volume Support (VS). Its proposed advantages
are to provide the positive attributes of
PSV with a constant minute volume. This is
closed-loop control of PSV, wherein tidal
volume provides feedback control for
continuously adjusting the pressure support
level. All breaths are patient triggered,
pressure limited, and flow cycled. The
pressure support level varies
breath-to-breath to maintain a constant
tidal volume. The maximum pressure change is
< 3 cm H2O and can range from 0
cm H2O above PEEP to 5 cm H2O
below the high pressure alarm setting.
Considerable speculation, but little data,
suggests that VS will wean the patient from
pressure support as patient effort increases
and lung mechanics improve. If the pressure
level increases in an attempt to maintain
tidal volume in the patient with airflow
obstruction, auto-PEEP may result. In cases
of hyperpnea, as patient demand increases,
ventilator support will decrease. This may
be the opposite of the desired response.
Additionally, if the minimum tidal volume
chosen by the clinician exceeds the patient
demand, the patient may remain at that level
of support and weaning may be delayed.
Dual Control Breath-to-breath
? Pressure limited time cycled ventilation
This approach is available as
Pressure Regulated Volume Control (PRVC),
Auto-Flow (Drager Evita 4), and Volume
Control Plus (VC+). This approach provides
the positive attributes of PCV with a
constant minute volume. This mode is a form
of pressure limited, time cycled ventilation
that uses tidal volume as a feedback control
for continuously adjusting the pressure
limit. All breaths are ventilator or patient
triggered, pressure limited, and time
cycled. The pressure increases or decreases
by ? 3 cm H2O per breath to
deliver the desired tidal volume. The
pressure limit fluctuates between PEEP and 5
cm H2O below the upper pressure
alarm setting. The proposed advantage of
PRVC is that it maintains the minimum peak
pressure that provides a constant set tidal
volume and automatic weaning of the pressure
as the patient improves. Perhaps the
greatest advantage of this mode is the
ability of the ventilator to change
inspiratory flow to meet patient demand
while maintaining a constant minute volume.
PRVC and similar modes are attractive with
implementing lung protective strategies
(such as the ARDSnet protocol), because the
tidal volume can me set to 6 mL/kg and the
peak pressure can be set to 30 cm H2O.
However, only anecdotal support of this
approach is currently available.
Conclusions
New ventilator modes and
related features have become available over
the past decade, with the claim that they
improve the efficiency and safety of
mechanical ventilation. Some also claim that
these modes facilitate the weaning process.
The decision to apply a particular mode of
ventilation, however, should also be based
upon an understanding of the underlying
physiology. Just because a new mode does
what it claims does not mean it will be more
useful than existing modes. Unfortunately,
there is very few clinical outcomes data
upon which to base a decision regarding the
choice of ventilator mode. The choice of a
particular mode is often based on clinician
experience and bias, institutional
preferences, and the capabilities of the
ventilators available at that institution.
References
Bonetto C, Calo MN, Delgado
MO, Mancebo J. Modes of pressure delivery
and patient-ventilator interaction. Respir
Care Clin N Am 2005; 11:247.
Branson RD, Davis K. Dual
control modes. Combining volume and pressure
breaths. Respir Care Clin NA 2001; 7:397.
Branson RD, Johannigman JA.
The role of ventilator graphics when setting
dual-control modes. Respir Care 2005;
50:187.
Branson RD, Johannigman JA.
What is the evidence base for the newer
ventilation modes? Respir Care 2004;
49:742-760.
Branson RD. Techniques for
automated feedback control of mechanical
ventilation. Semin Respir Crit Care Med
2000;21:203.
Du HL, Yamada Y. Expiratory
asynchrony. Respir Care Clin N Am 2005;
11:265.
Guttman J, et al. Automatic
tube compensation.
Respir Care Clin NA 2001;
7:475.
Habashi
NM.
Other approaches to open-lung
ventilation: airway pressure release
ventilation.
Crit Care Med
2005;
33(Suppl 3):
S228.
Hess D, Branson RD. New modes
of ventilation. IN: Hill NS, Levy MM.
Ventilator management strategies for
critical care. Marcel Dekker, 2001.
Hess D, Branson RD.
Ventilators and weaning modes. Respir Care
Clin NA 2000; 6:407.
Hess DR. Mechanical
ventilation strategies: what's new and
what's worth keeping? Respir Care 2002;
47:1007.
|